Oscillometry Healthcare Provider Toolkit

What is respiratory oscillometry? Definition, clinical applications and interpretation overview for healthcare professionals.

This toolkit is designed to help healthcare professionals better understand respiratory oscillometry, also referred to as forced oscillation technique (FOT).  Impulse oscillometry (IOS) is a sub-technique of respiratory oscillometry.  You can review the sections below on clinical perspectives and mechanics, as well as various tools and attachments. For more details on oscillometry and citations, please refer to the downloadable toolkit.  While this toolkit does not go into the details of oscillometry interpretation, it does provide a brief overview. 

Oscillometry Basics

Respiratory oscillometry is an objective lung function test that provides different information about the respiratory system that complements traditional lung function tests like spirometry and body plethysmography. Oscillometry is a non-invasive method for assessing the respiratory mechanics of the lung tissue (elasticity/stiffness), airways, and chest wall (compliance) during normal tidal breathing. 

An oscillometry device generates sound waves, which are superimposed on a patient's normal tidal breathing, to measure respiratory resistance and reactance (elastance or stiffness).  

Graphic of how oscillometry works, showing the input of oscillating sound waves at the mouth and how then they measure airway resistance

This effort-independent lung function test is different from spirometry. Oscillometry requires:

  • Passive breathing only – no forced maneuvers
  • Minimal cooperation or coordination from the patient

Therefore, it is useful for young children, those who cannot perform acceptable spirometry maneuvers and, in many situations, where spirometry is contraindicated.

Clinical Perspectives

How Does Oscillometry Work?

Oscillometry applies different frequencies of airflow or pressure, through sound waves, at the mouth during normal breathing. These sound waves are usually taken within a frequency range of 5 to 40 Hertz, which is higher than the frequency of normal breathing. The device then measures the oscillating pressure or flow generated by the lungs in response.  

Oscillometers used different techniques to generate these waves. The forced oscillation technique (FOT) delivers multi-frequency sinusoidal waves whereas the impulse oscillometry systems (IOS) deliver square waves as trains of individual frequencies. 

Oscillometry measures impedance (total opposition to airflow), which has components:

  1. Resistance (R): The degree of airway obstruction in the lungs (aka openness of airways)
  2. Reactance (X): The compliance (stiffness/elastance) of the lung tissue (aka expandability of airways)
What Are Key Terms in Oscillometry?

Impedance » Resistance + Reactance

  • Impedance: is the total opposition to airflow
  • Resistance (R): The degree of airway obstruction in the lungs
  • Reactance (X):  The compliance (stiffness/elastance) of the lung tissue.  How much the lungs can stretch and recoil.  

It is the resistance to airflow through the lung and is composed of:

  • Airway Resistance – Resistance from air moving through the airways
  • Tissue Resistance (tissue viscous resistance) – Opposition from the elastic recoil and deformation of lung and chest wall tissues

In most situations, the airways are the major contributors to the respiratory resistance such that R is a measurement of airway caliber. 

  • Compliance: The elasticity of the lung tissue.  How much the lungs can stretch and recoil.        By convention, this is expressed in negative terms.
  • Inertia: The mass of the air in the airways and is expressed in positive terms – as it is the energy expended by the pulse waves as it travels along the airways.  
  • Airway Diameters: a small change in radius of an airway significantly affects resistance (degree of airway obstruction)
  • Airflow Velocity/Turbulence: Turbulent flow (larger airways, high speed) creates more resistance than smooth laminar (smooth) flow (smaller airways, lower speed)
  • Air Viscosity and Length: Ticker air (viscosity) and longer airways increase resistance
  • Lung Volume/Chest Wall: Changes during breathing affect airway diameter and tissue elasticity.
  • Area of Reactance (AX): On the oscillometry test results, the area under the reactance curve from the lowest frequency to the resonant frequency (Fres).
  • Resonant Frequency (Fres): The frequency where Xrs equals zero.    
graph of the resistance chart
How Does Oscillometry Compare to Spirometry?

Oscillometry, like spirometry and plethysmography, are objective lung function tests.  This chart provides a comparison between these two objective lung function tests.   

Spirometry and Oscillometry: Evidence Based Comparison Chart

FeatureSpirometryOscillometry
What this objective test measuresbestLarge airway dynamicsPeripheral / small airway dynamics
Key measuresMeasures lung volumes and airflow:
  • FVC (Forced Vital Capacity)
  • FEV1 (Forced expiratory flow in 1st second)
  • Ratio for FEV1/FVC
Measures respiratory impedance using small amplitude oscillatory multi-frequency waves delivered at the mouth to measure respiratory:
  • Resistance
  • Reactance
Primary useGold-standard for the diagnosis of airflow obstruction (asthma, COPD)Highly sensitive to small-airway dysfunction, therefore useful across asthma, COPD, and other conditions for early detection or response to treatment
Level of patient effort requiredRequires forceful, maximal effort, often difficult for young children, elderly, or patient with illnessEffort-independent, requiring only tidal breathing – easier for young children and elderly and can be used in situations where spirometry is contra-indicated
Sensitivity to small-airway diseaseLimited; FEF 25-75 may correlate but inconsistentlySuperior sensitivity to small airway dysfunction, early detection of COPD symptoms and airway abnormalities
Measured dynamically or at restChallenges airways through forced expiratory flow which reveals abnormalitiesMeasures airways during normal breathing conditions and identifies abnormalities at rest
Bronchodilator response detectionChanges in FEV1/FVC used to assess responsivenessChanges in Resistance (R) and Reactance (X) used to assess responsiveness
More sensitive to detect bronchodilator responsiveness
Test durationLonger – requires repeated coached maneuvers and rest periodsShorter, quick to perform (30–60 second maneuvers)
Key metric to aid in medication therapyFEV1 guides medication therapy in asthma – NHLBI 2007
FEV1, along with patient’s symptoms and exacerbation risk, guide COPD therapy
No established metric to guide medication therapy
Patient burdenMay provoke cough, breathlessness, and fatigue (e.g. IPF)Associated with less symptom burden
Ideal populationsOlder children and adults who can follow instructionsAnyone but particularly very young children (>3 years), patients with cognitive or language barriers through lifespan
What Is the Clinical Usefulness of Oscillometry?
  • Be used across the lifespan
  • Assess young children
  • Monitor an individual’s environmental or occupational exposures over time
  • Provide insight into airway caliber
  • Assess reduced lung compliance
  • Detect small airway dysfunction
  • Track treatment response after bronchodilator or anti-inflammatory therapy
  • Predict COPD exacerbations
  • Manage disease long-term

Oscillometry may lead to improved health outcomes by:

  • Predicting COPD exacerbations. 
  • Informing treatment responsiveness and treatment plans during an asthma exacerbation 
  • Reactance parameters were found to be more sensitive in identifying poor asthma control than spirometry, supporting the use of oscillometry to complement spirometry in the clinical management of asthma
  • By assessing lung abnormalities early, oscillometry may then reduce the need for medications which could then reduce healthcare costs and the long-term effects of high-dose inhaled corticosteroids.  

The current gold standard test used to diagnosis asthma and COPD is spirometry. For this reason, most experts recommend that oscillometry should not be used as a standalone diagnostic tool but rather in conjunction with spirometry, plethysmography, and clinical history and other investigations such as chest imaging and appropriate blood work.

However, oscillometry can be clinically useful as outlined above in the “What Is the Clinical Usefulness of Oscillometry in the Assessment and Management of Lung Disease” section.

Indications for oscillometry use include:

  1. If the patient is unable to perform spirometry, oscillometry is recommended (i.e. too young, too old, or situations where patient is unable to follow instruction to perform spirometry; presence of contraindication to forced expiratory maneuvers but at risk for undiagnosed lung disease). Normal oscillometry can rule out underlying lung disease. Abnormal oscillometry indicates need for further assessment. 
  2. If results from spirometry are normal, but the patient remains symptomatic, oscillometry is recommended and subsequent follow up is indicated. 
  3. Oscillometry can be helpful in early detection and diagnosis of lung disease as it is more sensitive than spirometry at assessing small airway dysfunction.
  4. Oscillometry can be used to monitor medication treatment response.
  5. Monitor an individual’s environmental or occupational exposures over time. 
  6. Oscillometry can also help predict COPD exacerbations and can be used during an exacerbation to inform treatment. 
Considerations When Implementing an Oscillometry Test

When implementing oscillometry in clinical practice, consider the following:

  1. Have at least one clinic staff adequately trained in performing the procedure. 
  2. Establish written protocols and procedures to match the intended use and audience 
  3. Testing can be performed in an exam or procedure room. Some devices are portable and can be performed at bedside.
  4. Oscillometry should be conducted first, if other lung function tests such as spirometry are planned during the same clinic visit. This ensures that oscillometry is evaluated at FRC. Tests that require deep breaths and forced maneuvers will change the lung volume.
  5. Testing generally takes 15 minutes if it measures a patient’s baseline lung function.   If obtaining a pre- and post-bronchodilator response, 30 minutes should be allocated.

See “How to conduct an oscillometry test” section in the downloadable “Oscillometry:  A toolkit for healthcare professionals” for more information.  

Steps to conduct an oscillometry test include:

  1. Equipment and software set up
  2. Patient preparation
  3. Quality control during the test
  4. Determine acceptable measurements
  5. Determine measurement exclusion
  6. Ensure measurement repeatability
  7. Post-bronchodilator response – optional
  8. Reporting results

GRAPH PLACEMENT

Steps to Oscillometry Interpretation:

Please refer to the video above and the downloadable Oscillometry: A toolkit for healthcare professionals for further details on oscillometry test result interpretation.

Steps to interpretating oscillometry test results include:

  1. Check quality metrics (COV1 coherence)
    1. Coefficient of Variation: <10% for adults, <15% for children (R5 parameter)
    2. Repeat at least 3 tests for a valid session
  2. Review R5, total resistance
    1. R5 (resistance at 5Hz) reflects the total respiratory system resistance from central and peripheral airways
    2. Increased R5 suggests overall airway narrowing, as in asthma, COPD, or acute bronchoconstriction.
    3. Normal R5 means total airway resistance is within expected limits
  3. Review R5-R20, small airway resistance
    1. R20 (resistance at 20 Hz) primarily reflects large/central airways
    2. R5-R20 means small airway narrowing, often the earliest sign of obstructive lung disease
    3. A normal R20 but elevated R5-R20 can indicate isolated small airway disease
    4. R19 and R5-19 can be interpreted in the same way as R20 and R5-20
  4. Review X5:   Small airway reactance
    1. X5 (reactance at 5 Hz) reflects the elastic and inertial properties of peripheral airways
    2. More negative increased X5 values means stiffer, less compliant small airways
    3. This is seen in restriction, such as fibrosis or interstitial lung disease) and in severe small airway obstruction, such as asthma and COPD
  5. Review AX:   Area under the Reactance Curve
    1. AX is the integrated area under the reactance curve from 5 Hz to Fres (resonant frequency or zero-crossing)
    2. Increased AX indicates greater small airway dysfunction and can amplify findings from X5
    3. AX is highly sensitive to early changes in small airways
  6. Identify the pattern (peripheral, restrictive, obstructive)
  7. Add within-breath analysis, expiratory flow limitation or inspiratory flow limitation detection
  8. Consider reversibility, pre- and post-bronchodilator

According to technical standards supported by the American Thoracic Society (ATS) and European Respiratory Society (ERS), oscillometry (FOT) is used to detect asthma through measures of increased respiratory resistance (R) and reactance (X).   Recommended bronchodilator responsiveness (BDR) thresholds for asthma, based on the 95th percentile in healthy individuals, are defined as:

  • R5 ↓ by ≥40%
  • X5 ↑ by ≥50%
  • AX ↓ by ≥80%
  • Lack of reference equations
  • Lack of device and software standardization
  • Growing guideline support
  • Efficacy of clinical use in different populations and disease states
  • Need to define bronchodilator response thresholds
  • Provider familiarity with interpretation of results
  • Inability to bill for both oscillometry and spirometry at the same visit.  
  • Cost and equipment availability

The Mechanics of Oscillometry

How Does Oscillometry Use Sound Waves to Measure Lung Function?

In oscillometry, sound waves are artificially generated pressure oscillations that ride on top of a person’s normal breathing (tidal breathing). 

1. A loudspeaker (or a piston or a vibrating mesh) inside the oscillometry device is used to generate sound waves.

  • The speaker diaphragm moves back and forth.
  • This motion creates small pressure fluctuations in the air—basically sound waves.

2. How the oscillations enter the lungs

  • The patient breathes normally through a mouthpiece.
  • While the patient breathes, the device superimposes these sound waves into the airflow.
  • The oscillations are low-amplitude (gentle) and usually low frequency (about 5–35 Hz).

3. As the sound waves travel inside of the airways

  • Some energy is resisted by airway narrowing and friction.
  • Some energy is stored and released by the elastic tissues of the lungs.
  • Some energy reflects back due to changes in airway size.

This interaction changes the phase and amplitude of the sound waves.

4. Sensitive sensors measure:

  • Pressure changes
  • Flow changes

From these pressure and flow changes, the oscillometry device calculates respiratory impedance (total opposition), which includes:

  • Resistance (R) (airway obstruction) 
  • Reactance (X) (stiffness/elastance) 

5. Sound waves are perfect to be used for lung function measurement because they:

  • Travel easily through air
  • Can probe different airway sizes depending on frequency
  • Do not require forced breathing

By measuring how these waves change as they travel through the airways, the device can calculate respiratory impedance (total opposition), separating airway resistance (airway obstruction) and reactance (stiffness/elastance). Using different frequencies helps the system assess both large and small airways.

How to Conduct an Oscillometry Test
  1. Ensure that the patient is free of any active or suspected transmissible respiratory infection, such as coronavirus or tuberculosis.
  2. Ensure that the patient has not had any recent dental or facial surgeries, such as tooth extractions, and can form a proper tight seal around the mouthpiece.
  3. Ensure that the patient is as relaxed as possible, is not wearing tight-fitting clothing, and withholds from tobacco use and vigorous exercise at least 1 hour prior to testing.
  4. Perform oscillometry prior to other pulmonary function tests, such as spirometry.
  5. Ensure that the patient withholds bronchodilators prior to testing, unless instructed by a referring physician to continue bronchodilator medication. 
  1. Verify the resistance load of the oscillometry device by using a valid factory calibrated mechanical test load prior to patient testing.
  2. Remove the dust caps at both ends of the mechanical test load and attach them onto the oscillometry device.
  3. Upon successful verification, save and proceed with testing.
  4. Have multiple 'single-patient-use-bacterial/viral' filters and nose clips readily available.
  5. Have personal protection equipment (PPE), such as gloves and masks, and disinfectant wipes available.
  1. Verify the patient's information: first and last names, date of birth, birth sex, and height.
  2. Measure the patient's height without shoes, with feet together, standing as tall as possible with the eyes level and looking straight ahead, and the back flush against a wall or flat surface.
  3. Record patient's usage of bronchodilators, dosage, time/date of last administration, and any medication allergies.
  1. Ask the patient to sanitize their hands prior to entering the testing station.
  2. Outline the test duration of 30 seconds and the minimum requirement of three trials.
  3. Explain the sensation generated by oscillations such as 'vibrations' or 'fluttering.'
  4. Ensure that the patient is seated properly in a slight 'chin-up' position with both feet on the floor. Avoid slouching against the back of the chair or leg crossing.
  5. Instruct the patient to breathe normally while holding their cheeks with their palm and fingers and using their thumbs to support the soft tissue of the jaw during measurements. If they are unable to hold their cheeks themselves, you may hold it for them.
  6. Explain to the patient that swallowing should be avoided and the tongue must be below the mouthpiece during the test.
  1. Please refer to the manufacturer's instruction manual for individual instructions.
  2. Select New Patient, if appropriate, and enter the patient's information such as first and last names, date of birth, birth sex, height, weight, and smoking history.
  3. Ensure that the correct wavelength setup is selected. 
  4. Ensure the appropriate set of reference values is selected.  
  1. Click on Select Patient and choose the correct patient's file by verifying their information such as first and last names and date of birth.
  2. Ensure the patient's weight and height are updated prior to the start of testing.
  1. Attach a 'single-patient-use-bacterial/viral' filter to the oscillometry device.
  2. Ensure the oscillometry device is ready in the testing mode.
  3. Remind the patient of the 30 second test duration and the minimum requirement of three measurements.
  4. Instruct the patient to wear a nose clip.
  5. Adjust the oscillometry device to the patient's head level.
  6. Instruct the patient to wet their lips before wrapping them around the mouthpiece to form a proper, tight seal. Instruct the patient to begin to breathe normally.
  7. Observe the patient's breathing pattern and start recording following a minimum of three stable tidal breaths. Provide adequate rest time in between each measurement and adjust accordingly, based on the patient.
  1. Administer bronchodilator via a valued-holding chamber.
  2. Record the method and number of doses administered.
  3. Wait for at least 15 minutes.
  4. Repeat above testing steps to assess post-bronchodilator response

What is an acceptable oscillometry test?

  1. Ensure that the measurements have a validity greater than 70%.
  2. Inspect each measurement for anomalies or artefacts that may be caused by coughing, tongue obstruction, glottis closure, air leakage around the mouthpiece, attempting to talk, swallowing, and taking a deep breath.
  3. Exclude any unacceptable measurement with anomalies.
  4. Validity: Variation in consecutive R5 values < 15%

What is a repeatable oscillometry test?

  1. Ensure that a minimum of three acceptable measurements are recorded.
  2. The Coefficient of Variation (CoV) of Rrs for these measurements must be within the set limits:
    • Adults: CoV ≤ 10%.
    • Children: CoV ≤ 15%.
  1. Discard patient's mouthpiece and nose clip.
  2. Use disinfectant wipes to clean the oscillometry device and patient's chair.
  1. Include patient's first and last names, height, age, and birth sex.
  2. Include input signal frequencies and duration of individual recordings.
  3. Report the mean of acceptable and reproducible measurements and the CoV for these reported measurements.
  4. Select and report reference equations.
  5. Include impedance graph demonstrating Rrs and Xrs versus oscillation frequency.
  6. Include post-bronchodilator response with dosage and method of administration including z-scores and absolute percentage change - optional
What Factors Should Be Considered When Selecting an Oscillometer?

The following factors should be considered when selecting an oscillometer for a clinic setting:

  • FDA approval for intended clinical use 
  • Clinical validation data demonstrating accuracy, repeatability, and sensitivity 
  • Measurement parameters available (e.g., R5, R20, X5, AX, Fres) 
  • Quality control algorithms (coherence, leak detection, artifact rejection) 
  • Reference equations included and appropriateness for patient population 
  • Age range and feasibility (pediatric, adult, elderly) 
  • Ease of operation and consistency between operators 
  • Patient coaching aids (real‑time feedback, animations, visual cues) 
  • Turnaround time per test for busy clinic workflows 
  • Report clarity to aid clinical interpretability 
  • Data connectivity (electronic health record integration, export formats) 
  • Physical footprint and portability within clinic spaces 
  • Consumables and infection control compatibility 
  • Calibration procedures and ongoing maintenance needs 
  • Total cost of ownership (device, software, disposables, service) 
  • Vendor support, software updates
  • Training availability
How Is Oscillometry Coded for Legal and Appropriate Reimbursement?

Oscillometry is a reimbursable lung function test.  

CPT Code for Oscillometry

  • 94728: Airway resistance by impulse oscillometry.

Coding considerations

  • Bundling: Code 94728 cannot be reported with certain other pulmonary function tests, such as spirometry (94010) or bronchodilator response (94060).  If billed with another CPT code with a status indicator of S, T or V, it will be bundled (i.e. 94010, 94060, 94070, 94375, 94726).
  • Specificity: Always check the specific type of test performed to ensure the correct code is used, as there may be other codes for different types of airway resistance or lung volume measurements.
  • Modifiers: Modifiers may be needed in certain situations. For example, Modifier 26 (Professional Component used when billing only the provider’s professional work, but not the equipment, supplies, or technical side) and TC (Technical Component when the provider bills only the technical portion of a service and not the provider’s interpretation) can be used when the professional and technical components of the service are provided by different entities.

Resources

  1. Abdo, M., et al. (2023). Minimal clinically important differences in asthma outcomes. European Respiratory Journal, 61(3), 2201793. https://doi.org/10.1183/13993003.01793-202
  2. Berger, K. I., Wohlleber, M., Goldring, R. M., Reibman, J., Farfel, M. R., Friedman, S. M., … Shao, Y. (2021). Respiratory impedance measured using impulse oscillometry in a healthy urban population. ERJ Open Research, 7(1). https://doi.org/10.1183/23120541.00560-2020
  3. Bhattarai, P., et al. (2020). Clinical application of forced oscillation technique (FOT) in early detection of airway changes in smokers. Journal of Clinical Medicine, 9(9), 2778. https://doi.org/10.3390/jcm9092778 
  4. Biswas, R. (2025, July 16). How to interpret PulmoScan oscillometry data. PulmoScan. Retrieved Feb 1, 2026 from https://pulmoscan.cognitalabs.com/how-to-interpret-pulmoscan-oscillometry-data/
  5. Brown, N. J., Xuan, W., Salome, C. M., Berend, N., Hunter, M. L., Musk, A. W., James, A. L., & King, G. G. (2010). Reference equations for respiratory system resistance and reactance in adults. Respiratory Physiology & Neurobiology, 172(3), 162–168. https://doi.org/10.1016/j.resp.2010.05.013
  6. Brashier, B., & Salvi, S. (2015). Measuring lung function using sound waves: Role of the forced oscillation technique and impulse oscillometry system. Breathe, 11(1), 57–65. https://doi.org/10.1183/20734735.020514 
  7. Chang, E., Vasileva, A., Nohra, C., Ryan, C. M., Chow, C.-W., & Wu, J. K. Y. (2022). Conducting respiratory oscillometry in an outpatient setting. Journal of Visualized Experiments, (182), Article e63243. https://doi.org/10.3791/63243
  8. Children’s Colorado Pediatric Professionals. (2023, January 27). Measuring lung function with oscillometry [Video]. YouTube. https://www.youtube.com/watch?v=70C_etF84aA
  9. Corazalla, Edward O.  Personal communication.  January 28, 2026.
  10. Cottee, AM, Seccombe, LM, Thamrin C, King GG, Peters MJ, Farah CS.. (2020). Bronchodilator response assessed by the forced oscillation technique identifies poor asthma control with greater sensitivity than spirometry. Chest, 157(6), 1435–1441. https://doi.org/10.1016/j.chest.2019.12.035 
  11. Dellacà, R. L., et al. (2004). Detection of expiratory flow limitation in COPD using the forced oscillation technique. European Respiratory Journal, 23(2), 232–240. https://doi.org/10.1183/09031936.04.00046804 
  12. Desormeau, B., Smyrnova, A., Drouin, O., & Ducharme, F. M. (2023). Real-life impact of oscillometry in clinical assessment of preschoolers with asthma. Respiratory Medicine, 209, 107148. https://doi.org/10.1016/j.rmed.2023.107148
  13. Donohue, P. A., & Kaminsky, D. A. (2024). The role of oscillometry in asthma. Current Opinion in Pulmonary Medicine, 30(3), 268–275. https://doi.org/10.1097/MCP.0000000000001057 
  14. Ducharme, F. M., & Chan, R. (2025). Oscillometry in the diagnosis, assessment, and monitoring of asthma in children and adults. Annals of Allergy, Asthma & Immunology, 134, 135–143. https://doi.org/10.1016/j.anai.2024.11.013 
  15. Gochicoa-Rangel, L., & Vargas, M. H. (2025). How best to choose an oscillometer and reference equations for your patients with asthma. Annals of Allergy, Asthma & Immunology, 134, 159–164. https://www.annallergy.org/article/S1081-1206(24)01681-8/fulltext 
  16. Global Initiative for Chronic Obstructive Lung Disease. (2025). Global strategy for prevention, diagnosis and management of chronic obstructive pulmonary disease (2025 report). Retrieved from https://goldcopd.org/2025-gold-report/
  17. Gupta, N., Sachdev, A., Gupta, D., & Gupta, S. (2021). Oscillometry–The future of estimating pulmonary functions. Karnataka Pediatric Journal, 35(2), 79–87. https://doi.org/10.25259/KPJ_25_2020
  18. Hellinckx, J., De Boeck, K., Bande-Knops, J., et al. (1998). Bronchodilator response in 3–6.5‑year‑old healthy and stable asthmatic children. European Respiratory Journal, 12, 438–443. https://doi.org/10.1183/09031936.98.12020438 
  19. Hellyer, N., et al. (2017). Comparison of diaphragm thickness measurements among postures via ultrasound imaging. PM&R, 9(1), 21–25. https://doi.org/10.1016/j.pmrj.2016.06.001 
  20. Hickman, N., Hughes, A. L., Biswas, P., et al. (2023). Early detection of acute exacerbation of chronic obstructive pulmonary disease (AECOPD) using at‑home lung oscillometry. American Journal of Respiratory and Critical Care Medicine, 207(Suppl.), A4499. https://doi.org/10.1164/ajrccm-conference.2023.207.1_MeetingAbstracts.A4499 
  21. Huang, Y., Zhang, X., Wang, J., et al. (2025). Role of impulse oscillometry in chronic obstructive pulmonary disease and asthma–chronic obstructive pulmonary disease overlap. Clinical and Translational Allergy, e70057. https://doi.org/10.1002/clt2.70057 
  22. Kaminsky, D. A., Simpson, S. J., Berger, K. I., et al. (2022). Clinical significance and applications of oscillometry. European Respiratory Review, 31(163), 210208. https://doi.org/10.1183/16000617.0208-2021 
  23. King, G. G., Bates, J., Berger, K. I., et al. (2020). Technical standards for respiratory oscillometry. European Respiratory Journal, 55(2), 1900753. https://doi.org/10.1183/13993003.00753-2019 
  24. Krishnan, B., Kannukettiyil, N. J., & Gopal, A. (2025). Oscillometry versus spirometry: Initial insights. International Journal of Clinical Studies & Medical Case Reportshttps://ijclinmedcasereports.com/pdf/IJCMCR-RW-01278.pdf 
  25. Liang, X., Zheng, J., Gao, Y., et al. (2022). Clinical application of oscillometry in respiratory diseases: An impulse oscillometry registry. ERJ Open Research, 8(4), 00080‑2022. https://doi.org/10.1183/23120541.00080-2022 
  26. Malmberg, L. P., Pelkonen, A. S., Haahtela, T., et al. (2003). Exhaled nitric oxide rather than lung function distinguishes preschool children with probable asthma. Thorax, 58, 494–499. https://doi.org/10.1136/thorax.58.6.494
  27. Matos, R. I., McEvoy, C. T., & Jensen, E. A. (2025). Reversible airway obstruction on impulse oscillometry in preschool children with bronchopulmonary dysplasia. Pediatric Pulmonology. Advance online publication. https://pubmed.ncbi.nlm.nih.gov/41638357/
  28. Mondal P, Yirinec A, Midya V, et al. Diagnostic value of spirometry vs impulse oscillometry: A comparative study in children with sickle cell disease. Pediatric Pulmonology. 2019; 54: 1422-1430. https://doi.org/10.1002/ppul.24382
  29. Mou, T., Wang, Y., Fu, Y., et al. (2024). Analysis of the correlations and inconsistencies between spirometry and impulse oscillometry in the diagnosis of small‑airway dysfunction. BMC Pulmonary Medicine, 24, 619. https://doi.org/10.1186/s12890-024-03420-z 
  30. National Asthma Education and Prevention Program, Third Expert Panel on the Diagnosis and Management of Asthma. (2007). Expert Panel Report 3 (EPR-3): Guidelines for the diagnosis and management of asthma (NIH Publication No. 07-4051). National Heart, Lung, and Blood Institute, National Institutes of Health. https://www.nhlbi.nih.gov/health/pro/guidelines/current/asthma-guidelines
  31. Navanandan, N., Hamlington, K. L., Mistry, R. D., Szefler, S. J., & Liu, A. H. (2020). Oscillometry for acute asthma in the pediatric emergency department: A feasibility study. Annals of Allergy, Asthma & Immunology, 125(5), 607–609. https://doi.org/10.1016/j.anai.2020.07.002 
  32. Navanandan, N., et al. (2025). Clinical applications of impulse oscillometry in pediatric respiratory disease. Pediatric Pulmonology. Advance online publication. https://pubmed.ncbi.nlm.nih.gov/40705565/
  33. Nowowiejska, B., et al. (2008). Transient reference values for impulse oscillometry for children aged 3–18 years. Pediatric Pulmonology, 43(12), 1193–1197.
  34. O’Rourke, M. C., & Mendenhall, B. R. (2025). Transesophageal echocardiogram (TEE). StatPearls Publishing. https://www.ncbi.nlm.nih.gov/books/NBK430685/ 
  35. Oostveen, E., MacLeod, D., Lorino, H., et al. (2003). The forced oscillation technique in clinical practice: Methodology, recommendations and future developments. European Respiratory Journal, 22(6), 1026–1041. https://doi.org/10.1183/09031936.03.00089403 
  36. Oostveen, E., et al. (2013). Respiratory impedance in healthy subjects: Baseline values and bronchodilator response. European Respiratory Journal, 42(6), 1513–1523.
  37. Ostridge, K. (2019). The peak index: Spirometry metric for airflow obstruction severity and heterogeneity. Annals of the American Thoracic Society, 16(8), 974–975. https://doi.org/10.1513/AnnalsATS.201905-388ED 
  38. Park, H., Lee, H. J., Lee, H. W., et al. (2024). Diagnosis and evaluation of small airway disease and COPD using impulse oscillometry. Scientific Reports, 14, Article 79818. https://doi.org/10.1038/s41598-024-79818-w 
  39. Patel, S., Sylvester, K. P., Wu, Z., Rhamie, S., Dickel, P., Maher, T. M., Molyneaux, P. L., Calverley, P. M. A., & Man, W. D.-C. (2024). A comparison of respiratory oscillometry and spirometry in idiopathic pulmonary fibrosis: Performance time, symptom burden and test–retest reliability. ERJ Open Research, 10(4), 00227-2024. https://doi.org/10.1183/23120541.00227-2024
  40. Pisi, R., et al. (2021). Detection of small airway dysfunction in asymptomatic smokers with preserved spirometry: The value of the impulse oscillometry system. International Journal of Chronic Obstructive Pulmonary Disease, 16, 2585–2590. https://doi.org/10.2147/COPD.S319972 
  41. Rangel, L., Vargas, M. (2024). How best to choose an oscillometer and reference equations for your patients with asthma. Annals of Allergy, Asthma, and Immunology, 134 (2), 159-164. https://doi.org/10.1016/j.anai.2024.11.009 
  42. Ribeiro, C., Faria, A. C. D., Lopes, A. J., et al. (2018). Forced oscillation technique for early detection of the effects of smoking and COPD: Contribution of fractional order modeling. International Journal of Chronic Obstructive Pulmonary Disease, 13, 3281–3295. https://doi.org/10.2147/COPD.S173686
  43. Saadeh, C., Saadeh, C., Cross, B., Gaylor, M., & Griffith, M. (2015). Advantage of impulse oscillometry over spirometry to diagnose chronic obstructive pulmonary disease and monitor pulmonary responses to bronchodilators: An observational study. SAGE Open Medicine, 3, 2050312115578957. https://doi.org/10.1177/2050312115578957 
  44. Safia, N., Mayuran, R., Merin, K. (2022).  Asthma diagnosis using patient-reported outcome measures and objective diagnostic tests: Now and into the future. Current Opinion in Pulmonary Medicine, 28(3), 251-257. https://doi.org/10.1097/MCP.0000000000000871 
  45. Sarkar, S., Jadhav, U., Ghewade, B., Sarkar, S., & Wagh, P. (2023). Oscillometry in lung function assessment: A comprehensive review of current insights and challenges. Cureus, 15(10), e47935. https://doi.org/10.7759/cureus.47935
  46. Singh, D., Long, G., Cançado, J. E. D., & Higham, A. (2020). Small airway disease in chronic obstructive pulmonary disease: Insights and implications for the clinician. Current Opinion in Pulmonary Medicine, 26(2), 162–168. https://doi.org/10.1097/MCP.0000000000000637 
  47. Takahashi, S., Shirai, T., & Akamatsu, T. (2023). Oscillometry helps assess treatment responsiveness in adults with asthma exacerbations. Respiratory Physiology & Neurobiology, 313, 104065. https://doi.org/10.1016/j.resp.2023.104065 
  48. Thamrin, C., Gangell, C. L., Udomittipong, K., et al. (2007). Assessment of bronchodilator responsiveness in preschool children using forced oscillations. Thorax, 62, 814–819. https://doi.org/10.1136/thx.2006.071290 
  49. Usmani, O. S., et al. (2021). Why we should target small airways disease in our management of chronic obstructive pulmonary disease. Mayo Clinic Proceedings, 96(9), 2448–2463. https://doi.org/10.1016/j.mayocp.2021.03.016 
  50. Veneroni, C., Valach, C., Wouters, E. F. M., et al. (2024). Diagnostic potential of oscillometry: A population‑based approach. American Journal of Respiratory and Critical Care Medicine, 209(4), 444–453. https://doi.org/10.1164/rccm.202306-0975OC 
  51. Zimmermann, S. C., et al. (2020). Day‑to‑day variability of forced oscillatory mechanics for early detection of acute exacerbations in COPD. European Respiratory Journal, 56(3), 1901739. https://doi.org/10.1183/13993003.01739-2019 
  52. Zhang, Y., Tanabe, N., Shima, H., et al. (2022). Physiological impairments on respiratory oscillometry and future exacerbations in chronic obstructive pulmonary disease patients without a history of frequent exacerbations. COPD: Journal of Chronic Obstructive Pulmonary Disease, 19(1), 149–157. https://doi.org/10.1080/15412555.2022.2051005
  53. Zhu-Quan Su et al. on IOS in COPD and small airways (2018 submission, published 2018)
    Su, Z.-Q., Guan, W.-J., Li, S.-Y., Ding, M., Chen, Y., Jiang, M., Chen, X.-B., Zhong, C.-H., Tang, C.-L., Zhong, N.-S. (2018). Significances of spirometry and impulse oscillometry for detecting small airway disorders in COPD and heavy smokers. International Journal of Chronic Obstructive Pulmonary Disease, 13, 3031–3044. https://doi.org/10.2147/COPD.S172639 